Informace o publikaci

Externally Delivered Focused Ultrasound for Renal Denervation

Autoři

NEUZIL Petr ORMISTON John BRINTON Todd J. STÁREK Zdeněk ESLER Murray DAWOOD Omar ANDERSON Thomas L. GERTNER Michael WHITBOURNE Rob SCHMIEDER Roland E.

Rok publikování 2016
Druh Článek v odborném periodiku
Časopis / Zdroj JACC - Cardiovascular Interventions
Fakulta / Pracoviště MU

Lékařská fakulta

Citace
Doi http://dx.doi.org/10.1016/j.jcin.2016.04.013
Obor Kardiovaskulární nemoci včetně kardiochirurgie
Klíčová slova noninvasive; renal denervation; safety; treatment-resistant hypertension; ultrasound
Popis OBJECTIVES The aim of this study was to assess clinical safety and efficacy outcomes of renal denervation executed by an externally delivered, completely noninvasive focused therapeutic ultrasound device. BACKGROUND Renal denervation has emerged as a potential treatment approach for resistant hypertension. METHODS Sixty-nine subjects received renal denervation with externally delivered focused ultrasound via the Kona Medical Surround Sound System. This approach was investigated across 3 consecutive studies to optimize targeting, tracking, and dosing. In the third study, treatments were performed in a completely noninvasive way using duplex ultrasound image guidance to target the therapy. Short-and long-term safety and efficacy were evaluated through use of clinical assessments, magnetic resonance imaging scans prior to and 3 and 24 weeks after renal denervation, and, in cases in which a targeting catheter was used to facilitate targeting, fluoroscopic angiography with contrast. RESULTS All patients tolerated renal denervation using externally delivered focused ultrasound. Office blood pressure (BP) decreased by 24.6 +/- 27.6/9.0 +/- 15.0 mm Hg (from baseline BP of 180.0 +/- 18.5/97.7 +/- 13.7 mm Hg) in 69 patients after 6 months and 23.8 +/- 24.1/10.3 +/- 13.1 mm Hg in 64 patients with complete 1-year follow-up. The response rate (BP decrease >10 mm Hg) was 75% after 6 months and 77% after 1 year. The most common adverse event was post-treatment back pain, which was reported in 32 of 69 patients and resolved within 72 h in most cases. No intervention-related adverse events involving motor or sensory deficits were reported. Renal function was not altered, and vascular safety was established by magnetic resonance imaging (all patients), fluoroscopic angiography (n = 48), and optical coherence tomography (n = 5). CONCLUSIONS Using externally delivered focused ultrasound and noninvasive duplex ultrasound, image-guided targeting was associated with substantial BP reduction without any major safety signals. Further randomized, sham-controlled trials will be needed to validate this unique approach.

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info