Liquid Chromatography - Tandem Mass Spectrometry Method for Determination of Homocysteine in Rat Plasma: Application to the Study of a Rat Model for Tauopathies

Autoři

KOVAC Andrej ŠVIHLOVÁ Katarína MICHALICOVA Alena NOVAK Michal

Rok publikování 2014
Druh Článek v odborném periodiku
Časopis / Zdroj Journal of Chromatographic Science
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
Obor Analytická chemie, separace
Klíčová slova liquid chromatography; mass spectrometry; rat plasma; homocysteine; tauopathies
Popis Hyperhomocysteinemia is a common occurrence in many neurodegenerative diseases, including tauopathies. We developed and validated a simple and sensitive liquid chromatography-tandem mass spectrometry method for the analysis of homocysteine (Hcy) in rat plasma. The method showed excellent linearity with regression coefficients higher than 0.99. The accuracy was in the range of 93-98%. The inter-day precision (n=5 days), expressed as % relative standard deviation, was in the range 3-8%. The process was validated in compliance with the EMA quideline. Using this method, we analyzed plasma samples from two transgenic lines of the rat model for tauopathies.