Informace o publikaci

Low-Level Radon Activity Concentration—A MetroRADON International Intercomparison

Autoři

OTAHAL Petr P. S. FIALOVÁ Eliška VOSAHLIK Josef WIEDNER Hannah GROSSI Claudia VARGAS Arturo MICHIELSEN Nathalie TURTIAINEN Tuukka LUCA Aurelian WOŁOSZCZUK Katarzyna BECK Thomas R.

Rok publikování 2022
Druh Článek v odborném periodiku
Časopis / Zdroj International Journal of Environmental Research and Public Health
Fakulta / Pracoviště MU

Přírodovědecká fakulta

Citace
www https://www.mdpi.com/1660-4601/19/10/5810/pdf
Doi http://dx.doi.org/10.3390/ijerph19105810
Klíčová slova radon; interlaboratory comparison; radon activity concentration; calibration; traceability
Přiložené soubory
Popis An international comparison of continuous monitors measuring radon activity concentration was performed to validate the traceability of the European radon calibration facilities. It was carried out by comparing the secondary standards used by these previous facilities, ranging from 100 Bq·m-3 to 300 Bq·m-3. Secondary standards were individually compared to a secondary reference device previously calibrated in a reference radon atmosphere traceable to a primary standard. The intercomparison was organized by the National Institute for Nuclear, Chemical, and Biological Protection (SUJCHBO) in the period from October 2019 to April 2020 within the European Metrology Program for Innovation and Research (EMPIR), JRP-Contract 16ENV10 MetroRADON. Eight European laboratories participated in this study. The results of the experiment are presented and discussed.

Používáte starou verzi internetového prohlížeče. Doporučujeme aktualizovat Váš prohlížeč na nejnovější verzi.

Další info