Real-time PCR diagnostics failure caused by nucleotide variability withing exos 4 of human cytomegalovirus major immediate-early gene

Autoři

LENGEROVÁ Martina RÁČIL Zdeněk VOLFOVÁ Pavlína LOCHMANOVÁ Jana BERKOVCOVÁ Jitka DVOŘÁKOVÁ Dana VORLÍČEK Jiří MAYER Jiří

Rok publikování 2007
Druh Článek v odborném periodiku
Časopis / Zdroj Journal of Clinical Microbiology
Fakulta / Pracoviště MU

Lékařská fakulta

Citace
Obor Onkologie a hematologie
Popis Here we report how variability in the human cytomegalovirus genome sequence may seriously affect the outcome of its molecular diagnosis. A real-time quantitative PCR assay targeting the major immediate-early gene failed to detect the viral load in some, but not all, clinical samples from hematooncological patients. By amplification and sequencing the DNA across the regions targeted by this assay we found a number of nucleotide substitutions which accounted for decreased primer/probe binding. This decreased the sensitivity of the assay up to 1,000-fold.