Publication details

The determination of total protein in serum. Notes to the IFCC method, Biuret reaction mechanism

Authors

CHROMÝ Vratislav HLAVÁČOVÁ Barbora ŠVÁCHOVÁ Lenka

Year of publication 2007
Type Article in Periodical
Magazine / Source Clinical Chemistry and Laboratory Medicine
MU Faculty or unit

Faculty of Science

Citation
Field Biochemistry
Keywords Biuret method mechanism
Description The calibration of serum protein by bovine albumine determined by the biuret reaction leads to falsely higher protein results of about +5%.