Publication details

Ultra-trace mercury determination in drinking water by APGD-OES with cold vapor generation and gold trap enrichment

Authors

KLIŚ Tymoteusz GRĘDA Krzysztof ČECH Jan JAMRÓZ Piotr MEDEK Tomáš PAZDERKA Michal DVOŘÁK Pavel POHL Paweł

Year of publication 2027
Type Peer-reviewed scientific article
Magazine / Source Talanta
MU Faculty or unit

Faculty of Science

Citation
web https://www.sciencedirect.com/science/article/pii/S0039914026011628
Doi https://doi.org/10.1016/j.talanta.2026.130506
Keywords Water analysis; Atmospheric pressure glow discharge; Optical emission spectrometry; Gold amalgamation; Mercury preconcentration
Attached files
Description An atmospheric pressure glow discharge (APGD) optical emission spectrometry (OES) method for ultra-trace Hg determination in water was developed by combining cold vapor (CV) generation with a custom-built gold trap (GT). Among the two high-voltage power supplies evaluated (pulsed DC and constant DC), the constant DC-driven microplasma yielded superior Hg excitation. Under optimized conditions, limits of detection (LODs) for Hg in CV-APGD-OES were 0.2 and 0.08 µg L-1 for 1 s and 10 s integration times, respectively. Coupling with the GT unit improved sensitivity by over 100-fold, yielding an LOD of 0.003 µg L-1, well below regulatory limits for drinking water. Time-resolved signal measurements (intensity of the Hg I 253.7 nm line) were fitted to an exponentially modified Gaussian function (EMG), enabling repeatabilities of 4.2% and 2.2% at concentration levels of 0.25 and 5 µg L-1, respectively. The method was validated using spiked water samples (sub-µg L-1 levels), with recoveries of 98–108%. Compared to conventional, bulky systems, the proposed approach offers reduced costs, simplified instrumentation, and lower maintenance requirements.
Related projects: